Abstract
PURPOSE: The purpose of this study was to examine, in a controlled environment, effects of connection design upon screw stability. Implant fixtures have 2 types of connections to the abutment: internal connection and external hex. Four implant systems were tested: Bio-Lok (external hex; Bio-Lok International, Inc., Deerfield Beach, FL); Zimmer (internal connection; Zimmer Dental, Carlsbad, CA); Nobel Biocare (external hex; Nobel Biocare USA, Inc., Yorba Linda, CA); and Astra Tech (internal connection; Astra Tech Inc., Waltham, MA). MATERIALS AND METHODS: Ten samples of each system, including base, implant, abutment, and molar crown, were loaded to 200 N for 1 × 10 cycles. Screws were tightened to manufacturers' recommendations,and torque audits done at 2.5 × 10, 5 × 10, 7.5 × 10, and 1 × 10 cycles. RESULTS: The Bio-Lok samples lost an average of 10% of the original torque values, the Astra Tech group lost almost all of the torque and loosened, while the Zimmer and Nobel Biocare samples lost an average of 50% of the torque but did not loosen (P ≤ 0.05). CONCLUSIONS: It may be concluded from this study that although internal connections are clinically favored, this study did not show any advantage relative to screw loosening. However, screw design may be a significant factor in loosening of the joint.
| Original language | English |
|---|---|
| Pages (from-to) | 427-435 |
| Number of pages | 9 |
| Journal | Implant Dentistry |
| Volume | 15 |
| Issue number | 4 |
| DOIs | |
| State | Published - 2006 |
| Externally published | Yes |
ASJC Scopus Subject Areas
- Oral Surgery
Keywords
- External hex
- Implant
- Internal connection
- Screw loosening
Fingerprint
Dive into the research topics of 'An in vitro analysis of implant screw torque loss with external hex and internal connection implant systems'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS