Skip to main navigation Skip to search Skip to main content

Finite Element Analysis of an Implant-Supported FDP with Different Connector Heights

  • Laura H. J. Alberto
  • , Lohitha Kalluri
  • , Josephine F. Esquivel-Upshaw
  • , Yuanyuan Duan

Research output: Contribution to journal › Article › peer-review

Abstract

All-ceramic fixed dental prostheses (FDPs) tend to fracture in the connector areas, due to the concentration of tensile stresses. This study aimed to evaluate the role of connector height on the stress distribution of a posterior three-unit implant-supported all-ceramic FDP using finite element analysis (FEA). Two titanium dental implants, their abutments, screws, and a three-unit all-ceramic FDP were scanned using a micro-CT scanner. Three 3D models with altered distal connector heights (3, 4, and 5 mm) were generated and analyzed on ABAQUS FEA software. The maximum principal stress values in MPa observed for each model with different connector heights and their respective locations (MA = mesial abutment; DA = distal abutment; F = framework; V = veneer) were: 3 mm—219 (MA), 88 (DA), 11 (F), 16 (V); 4 mm—194 (MA), 82 (DA), 8 (F), 18 (V); 5 mm—194 (MA), 80 (DA), 8 (F), and 18 (V). All the assembled models demonstrated the peak stresses at the neck area on the mesial abutments. The connector height had a significant influence on the stress distribution of the prosthesis. The models with higher distal connectors (4 and 5 mm) had a lower and more uniform distribution of maximum principal stresses (except for the veneer layer) when compared with the model with the smallest distal connector.
Original languageEnglish
Article number2334
JournalSymmetry
Volume14
Issue number11
DOIs
StatePublished - Nov 7 2022
Externally publishedYes

Funding

This study was supported by NIH-NIDCR grant R56 DE025001 and the University of Florida Clinical and Translational Science Institute, which is supported in part by the NIH National Center for Advancing Translational Sciences under award number UL1TR001427. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. Ceramic materials were provided by Ivoclar Vivadent and implant materials were provided by Dentsply. This study is registered at ClinicalTrials.gov under study number NCT01729858.

Keywords

  • all-ceramic FDP
  • connector design
  • connector height
  • dental biomaterials
  • dental implants
  • design parameters
  • finite element analysis

Fingerprint

Dive into the research topics of 'Finite Element Analysis of an Implant-Supported FDP with Different Connector Heights'. Together they form a unique fingerprint.

Cite this