Abstract
Background and Purpose
Sequential compression is used to manage lymphedema, but little is known about pressures delivered to the therapeutic targets. This study characterized actual pressures delivered by a traditional compression pump (Lympha Press [LP]) and one using an alternate compression pattern (Flexitouch [FT]).
Subjects
Ten adults who were healthy volunteered to participate in the study.
Methods
Pressure-time along the forearm was measured using a 256–pressure sensor array during the pressure cycling of each device. Device assessments were separated by at least 48 hours.
Results
Pressure patterns and magnitudes produced by the 2 devices differed considerably. The FT pressure pattern displayed a rapid rise and fall, progressing from the wrist toward the elbow. The LP pressure rose slower and was sustained at a higher level during its inflation cycle. Pressures delivered with the LP were significantly greater than those delivered with the FT.
Discussion and Conclusion
The pressure patterns and magnitudes on treated limbs depend on the device. These differences should be considered before selecting a device for a specific patient.
Sequential compression is used to manage lymphedema, but little is known about pressures delivered to the therapeutic targets. This study characterized actual pressures delivered by a traditional compression pump (Lympha Press [LP]) and one using an alternate compression pattern (Flexitouch [FT]).
Subjects
Ten adults who were healthy volunteered to participate in the study.
Methods
Pressure-time along the forearm was measured using a 256–pressure sensor array during the pressure cycling of each device. Device assessments were separated by at least 48 hours.
Results
Pressure patterns and magnitudes produced by the 2 devices differed considerably. The FT pressure pattern displayed a rapid rise and fall, progressing from the wrist toward the elbow. The LP pressure rose slower and was sustained at a higher level during its inflation cycle. Pressures delivered with the LP were significantly greater than those delivered with the FT.
Discussion and Conclusion
The pressure patterns and magnitudes on treated limbs depend on the device. These differences should be considered before selecting a device for a specific patient.
| Original language | English |
|---|---|
| Pages (from-to) | 1379-1388 |
| Number of pages | 10 |
| Journal | Physical Therapy |
| Volume | 87 |
| Issue number | 10 |
| DOIs | |
| State | Published - Oct 2007 |
ASJC Scopus Subject Areas
- Physical Therapy, Sports Therapy and Rehabilitation
Disciplines
- Physical Therapy
Fingerprint
Dive into the research topics of 'Interface pressures produced by two different types of lymphedema therapy devices'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS